2. Put the other end of the hose into a drain bucket.
3. Insert part (5) into valve (3) and press it until you hear a click. The drain
valve opens immediately.
4. When the water reservoir is empty, press part (4). Item (5) will pop out
approx. 1/8"(3mm) and the drain valve will be closed. Remove item (6) with
the plastic tube.
Caution!
Never reuse waste water.
5. If the drained reservoir is the clean-water reservoir, fill reservoir with distilled
water until it reaches the full level. (Approximately 219 Ounces /6.5 liters).
The autoclave is now ready for use.
8.12.1
Waste water draining
Caution!
Waste water should be brought into the public net in accordance with
the local rules or requirements i.e ONLY NON-HAZARDOUS LIQUIDS
SHALL BE DISPOSED IN PUBLIC SEWAGE!
Connect the autoclave's drain to the building’s drainage pipe. The drainage
shall be of an open type, withstanding temperature of, at least, 140ºF.
(Applicable for device with automatic water inlet system only).