Message
Description
Please fill water tank to full
for start
For prion cycle in manual water filling
configuration, when the water tank is not
full the “Start” button will not appear,
instead this error message will appear in
the Active alarms screen.
Prion cycle, in manual water filling and
the water tank is not full
Unrecognized printer
If printer is not recognized an error will
appear in the UI.
Low Quality Water
1. Read the water level from analog
sensor.
2. Translate the digital value to micro
siemens according to the attached
table.
3. Compare
the
value
system
parameter
(water
quality
reference, default 10).
4. if value is less than 10 show error -
Low Quality Water
Set Atmosphere Pressure
is Active
Present when Reset Atmospheric
pressure is requested from Technician
screen
“Not enough water
detected”
If in the clean water tank, the Float switch
does not change from 1 to 0 during water
pump time On
Pipe water heater
temperature time error
If the Pipe Water Heater works for more
than 15 min.
Jacket water heater
temperature time error
If the Jacket Water Heater works for more
than 15 min.
“Analog Input Error - Jacket
Temperature”
Page 75
Содержание AMS11-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Страница 33: ...2 Select the Version information 3 The Version information is displayed For an example see below Page 26...
Страница 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Страница 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Страница 86: ...Page 79...
Страница 105: ...Cat No MAN205 0502003EN Rev C COPYRIGHT...