2.2 Safety features incorporated in the device
The pressure vessel chamber door has the following features protecting
personnel from hazards:
l
Two door switches indicate that the door is closed. Without this indication
steam is not introduced into the chamber.
l
An electrical door locking pin that blocks door opening during operation.
The following safety devices are installed in the autoclave to optimize its safe
operation:
l
Safety thermostat, to prevent over- heating of the chamber heating
elements.
l
Safety cut-off switch to prevent over-heating of the pipe heating element.
l
Safety pressure valve to prevent over-pressurizing of the chamber.
Warning!
Mind the Power socket. Keep it and its vicinity dry.
Danger of electrocution.
Page 17
Содержание AMS11-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Страница 33: ...2 Select the Version information 3 The Version information is displayed For an example see below Page 26...
Страница 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Страница 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Страница 86: ...Page 79...
Страница 105: ...Cat No MAN205 0502003EN Rev C COPYRIGHT...