Substance
Feed Water
Condensate
Chloride (Cl)
≤ 2 mg/l
≤ 0.1 mg/l
Phosphate
≤ 0.5 mg/l
≤ 0.1 mg/l
Conductivity (at
20°C)
≤15 µs/cm
≤ 3 µs/cm
pH value
5 to 7.5
5 to 7
Hardness
≤ 0.02 mmol/l
≤ 0.02 mmol/l
Appearance
Colorless, clean, without sediments
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Caution:
The use of water for autoclaves that do not comply with the table
above may have severe impact on the working life of the sterilizer
and can invalidate the manufacturer’s guarantee.
Use only deionized water, having a maximum conductivity of 15
µs/cm. Conductivity greater than 15 µs/cm may cause failures.
1.12.1
Tap water supply
The range of hardness value 0.7-2.0 mmol/l (70- 200 mg/l CaCO3)
The use of soft water is strictly forbidden!
Please consult a water specialist!
Note:
We recommend testing the water quality once a month. The use of water
for autoclaves that does not comply with the table above may have severe
impact on the working life of the sterilizer and can invalidate the manufacturer’s
guarantee.
Page 14
Содержание AMS11-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Страница 33: ...2 Select the Version information 3 The Version information is displayed For an example see below Page 26...
Страница 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Страница 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Страница 86: ...Page 79...
Страница 105: ...Cat No MAN205 0502003EN Rev C COPYRIGHT...