5.6 Before starting the Autoclave
Important!
Reset the Atmospheric Pressure prior to operating the Autoclave for the
first time.
To reset the Atmospheric Pressure:
1. Open the Autoclave door and verify that the Chamber's ambient
temperature is below 113°F.
2. In the
Quick Option
screen, press the
Settings
icon
, then browse
to Maintenance/Reset atmospheric pressure.
3. In the confirmation screen, press
Reset
.
Before starting the autoclave we recommend that you check the following:
1. In the
Quick Option
screen, press the
Info
option.