66
Req ID
Message
Description
Msg-130
"Chamber
temperature not
in range"
This message is displayed if the temperature in the
chamber > End temperature (120C) or < 2C
Msg-140
"Chamber
pressure not in
range"
This message is displayed if Chamber pressure <ATM
pressure-40kPa or >ATM Pr10kPa).
Msg-150
Canceled.
Msg-160
"I/O card is not
connected"
This message is displayed if I/O card is disconnected (both
while cycle is running or not).
Msg-170
"Low Temp"
This message is displayed if the temperature drops for
more than 1 second below the sterilization temperature
during sterilization cycle.
Msg-180
"High Temp"
This message is displayed if the temperature raises 3°C
above sterilization temperature during the sterilization
stage for 2 seconds during sterilization cycle.
Msg-190
"High Temp.
(Ending)"
This message is displayed if the system cannot reach the
required temperature, in the chamber, within 10 minutes.
Msg-200
"Heat Time Error"
This message is displayed if the system cannot reach the
required temperature, in the chamber, within the preset
time.
Msg-210
"Low Pressure"
This message is displayed if Chamber Pressure drops
below the sterilization pressure for 2 seconds during the
sterilization stage.
Msg-220
"High Pressure"
This message is displayed if Chamber Pressure raises 29
kPa above sterilization pressure for 2 seconds during the
sterilization stage.
Msg-230
"High Pressure
(Ending)"
This message is displayed if the system cannot reach
atmospheric pressure ± 5kPa during the ending stage.
Msg-240
"High Pressure
(Exhaust)"
This message is displayed if the system cannot reach
preset pressure within 10 minutes from the beginning of
the exhaust stage.
Msg-250
Canceled.
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...