Launder textile wraps prior to sterilization, thoroughly rinse wraps laundered in
chlorine bleach. Chlorine bleach can harm your stainless-steel instrument and the
sterilizer.
Follow the instrument manufacturer’s instructions on the use of products for
cleaning and lubricating instrument that have been ultrasonically cleaned.
Be sure that instruments of dissimilar metal (stainless steel, carbon steel, etc.) are
separated. Carbon steel instruments should be bagged or placed on autoclavable
towels and not directly on stainless steel trays (mixing will result in damage to the
instruments or trays from the oxidation of these materials).
Load items within the boundaries of the tray so that they do not touch the chamber
walls or fall off when the tray is moved. Items should not be allowed to touch the
walls of the Chamber as the hot metal can damage the item.
Don’t overload the Sterilizer trays (see Specification). Overloading will cause
inadequate sterilization & drying.
Make sure that all instruments remain apart during the sterilization cycle. Surfaces
that are hidden because items are covering other items will not be exposed to the
steam and will not be sterilized.
Disassemble or sufficiently loosen multiple-part instruments prior to packaging to
permit the sterilizing agent to come into direct contact with all parts of the
instrument.
Verify that packaging methods are in accordance with the good practice approach
and the packaging materials used are in agreement with applicable standards.0
Tilt on edge items prone to entrap air and moisture, e.g. hollowware, so that only
minimal resistance to air removal exist, the passage of steam and condensate will
be met.
Wrapped instruments should be placed in material which will allow steam
penetration and promote drying, such as autoclave bag, autoclave paper, or muslin
towels.
When loading pouches on the tray, put them with paper side up, nylon side towards
the tray (see the figure below)
Tubing should be rinsed after cleaning. When placed in the tray, make sure that
both ends of the tubing are open and there are no sharp bends or twists.
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...