56
To drain the reservoir:
1. See item (5) with the plastic hose (6) attached to it (supplied with the autoclave).
2. Put the other end of the hose into a drain bucket.
3. Insert part (5) into valve (3) and press it until you hear a click. The drain valve opens
immediately.
4. When the water reservoir is empty, press part (4). Item (5) will pop out approx. 3mm
and the drain valve will be closed. Remove item (6) with the plastic tube.
Caution!
Never reuse waste water.
5. If the drained reservoir is the clean-water reservoir, fill reservoir with distilled water
until it reaches the full level. (Approximately 6.5 liters).
The autoclave is now ready for use.
8.9.1
Waste water draining
Caution!
Waste water should be brought into the public net in accordance with the local
rules or requirements i.e ONLY NON-HAZARDOUS LIQUIDS SHALL BE
DISPOSED IN PUBLIC SEWAGE!
Connect
the
autoclave's drain to the building’s drainage pipe. The drainage shall be of an open
type, withstanding temperature of, at least, 60
º
C. (Applicable for device with automatic water
inlet system only).
Waste Water
Clean Water
3
4
5
6
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...