68
Req ID
Message
Description
Msg-390
Water tank filling,
Please wait
If high water demand cycle is chosen (e.g. Prion) and
Parameter>0 (Auto fill) - Then before enabling Start
message presented on GUI
Msg-400
Please fill water
tank to full for
start
For prion cycle in manual water filling configuration, when
the water tank is not full the “Start” button will not
appear, instead this error message will appear in the
Active alarms screen.
Prion cycle, in manual water filling and the water tank is
not full
Msg-410
Unrecognized
printer
If printer is not recognized an error will appear in the UI.
Msg-420
Low Quality
Water
1.
Read the water level from analog sensor.
2.
Translate the digital value to micro siemens
according to the attached table.
3.
Compare the value system parameter (water
quality reference, default 10).
4.
if value is less than 10 show error - Low Quality
Water
Msg-430
Set Atmosphere
Pressure is Active
Present when Reset Atmospheric pressure is requested
from Technician screen
Msg-440
“No Water”
Chamber clean cycle, If Float switch (DI 2) doesn’t change
from 1 to 0 during water pump time On.
Msg-450
“Analog Input
Error - Jacket
Temperature”
Msg-460
Utility issue #1 –
Please switch OFF
and ON
machine’s power
switch
If windows card disconnects
Msg-470
Utility issue #2 –
Please switch OFF
and ON
machine’s power
switch
If WIN application crash/not start up
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...