14. Please fill water tank to full for start.Chamber Clean Program-Alarm
water fill before start,
15. Chamber clean program Stage Cleaning-insert water
The chamber clean process last 45 minutes
The warming up to 80ºC and hold for 10 minutes at 80ºC, (This stage uses
500ml of water from the clean water reservoir) .Exhaust water from the
chamber, goes to air removal , pass to plateau of 3 minutes at 132 ºC and
1 minute drying time.