60
2. Disconnect the pipe.
3. Connect the new filter.
4. Screw in the new filter.
5. Ensure the new filter is all the way in and seated properly.
Note:
make sure that the arrow on the filter body points inwards, toward the
chamber. Make sure that you don’t bend the filter pipe when reattaching it.
Note:
It is recommended to replace the air filter every 6 months or after 1000
cycles (whichever is the shorter period).