67
Req ID
Message
Description
Msg-260
"Pressure Time
Error"
This message is displayed if the system cannot reach the
required pressure conditions in the chamber, after preset
time, during the air removal stage.
Msg-270
N/A
N/A
Msg-280
"Time Error"
This message is displayed if the real time clock is faulty.
Msg-290
"Door is open
(During the
cycle)"
This message is displayed when the door is open: During
the cycle.
Msg-300
"Canceled By
User"
This message is displayed after the STOP button is pressed
and cycle aborted.
Msg-310
This message and symbol are displayed if an error occurs
before sterilization cycle is completed.
Msg-320
"Air Error"
This message is displayed at the end of the cycle if the
autoclave does not reach the atmospheric pressure after
10 minutes.
Msg-330
Canceled.
Msg-340
If the water level is low in the clean water reservoir and
not sufficient for at least one cycle, this symbol will be
displayed.
Mineral free water reservoir empty will appear in the
error messages.
Msg-350
"Routine cycle
service is
recommended
Please call your
service provider."
Number of cycles, since last periodical maintenance,
exceeded the "cycle service counter" parameter Or time
elapsed exceeded the "time service counter" parameter.
Msg-360
"Power Down"
This message is displayed if power down has occurred
during the cycle. (this message will print out in the printer
after the autoclave will turn on).
Msg-370
Vacuum time
error”
Vacuum pressure fails to reach the required value within
the required time (in Vacuum test and Pre-vacuum stage)
Msg-380
Water fill Error
If DO10 is On for timer > Parameter Water fill time out
and DI2 (Max. float switch) is not On
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...