Page 14
3.8 Overall Dimensions
3.9 Electrical Data
Property
Value
Total Power
3000W
Voltage
230VAC / 1 ph
Amperage
10A
Protection against electrical
shock
Class I (IEC 60601-1)
Mains supply fluctuation
+/- 10%
Degree of protection by
enclosure
IP31
Note:
In order to avoid any injury by electrical hazard, it is recommended that
a ground fault protection device be installed in the electrical panel
feeding the autoclave (local codes may make this mandatory).