Page 76
11.7 Maintenance
Maintenance procedures provided by Bacsoft software allow you
additional tests and USB input/output options.
Browse to the following folder:
Main menu\Maintenance
You will see the following screen listing the maintenance options:
Below is the instruction for autoclave’s maintenance menu.
11.7.1 Reset atmospheric pressure
In this menu you can reset the atmospheric pressure value. To do so:
1.
Browse to the following folder:
Maintenance\Reset atmospheric pressure
The following screen will appear:
2.
Leave the door open for 2 minutes at least. Ambient temperature
should be less than 45°C.
Note:
Please reset the atmospheric pressure when you install the autoclave
for the first time, and each time you relocate or calibrate the
autoclave.