Page 38
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.10 Program 10: Bio Hazard 2 (BH, bio-hazard models
only)
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter.
For
instruments, when the manufacturer recommends autoclaving at
temperatures of 121
C (250
F) for 45 minutes. Drying stage is
available for PV (pre-vacuum) models only.
Nominal parameters default settings
Sterilization temperature: 121
C (250
F)
Sterilization time: 45 minutes
Drying time: 1 minute. (PV – pre-vacuum models only).
Operations sequence:
(PV, pre-vacuum models only) Pulse low/Pulse high:
at one
pulse it will build vacuum down to 25 kPa.
All exhaust from the
chamber is performed through the bio-hazard filter.
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
Cooling:
N/A
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Note:
If the cycle fails, fast exhaust is performed through the bio-hazard filter.
Some water may remain in the chamber after a failed cycle. To remove
such water, press Start and repeat the cycle. If the problem persists,
call for service.
(PV, pre-vacuum models only) Drying:
vacuum is built up and the
air pump is working.