Page 39
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.11 Program 11: Bio Hazard Liquids (BH, bio-hazard
models only)
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter.
For
liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250
F) for 45 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
Sterilization temperature: 121
C (250
F)
Sterilization time: 45 minutes
Operations sequence:
(PV, pre-vacuum models only) Pulse low/Pulse high
: at one
pulse it will build vacuum down to 25 kPa.
All exhaust from the
chamber is performed through the bio-hazard filter.
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
Cooling:
N/A