5.7 Program 7: Waste (PV – pre-vacuum models only)
For waste products when the manufacturer recommends autoclaving at
temperatures of 134
C (273
F).
Nominal parameters default settings
Sterilization temperature: 134
C (273
F)
Sterilization time: 7 minutes
Drying time: 15 minutes.
Operations sequence:
Pulse low/Pulse high:
four vacuum pulses remove air by drawing
the chamber down to 25kPa.
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
Cooling:
N/A
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Drying:
vacuum is built up and the air pump is working.