Page 42
5.14 Program 14: Warm-Up
Pre-warming the chamber at 80
C (176
F) without drying.
Caution!
This is not a sterilization program!
Nominal parameters default settings
Warm-up temperature: 80
C (176
F).
Warm-up time: 20 minutes.
Operations sequence:
Pulse low/Pulse high:
N/A
Heating:
Steam enters the chamber and heats it up until the warm-
up temperature is reached.
Warm up:
Warm up temperature is maintained constant for the
warm-up time.
Sterilization:
N/A.
Cooling:
N/A.
Fast Exhaust:
Steam is exhausted out of the chamber at a fast rate
until pressure decreases to ambient pressure.
Drying:
N/A.
Note:
Some water may remain in the chamber.