5.13 Program 13: Bowie and Dick Test (PV – pre-vacuum
models only)
Bowie and Dick test is a test program with the following parameters:
Nominal parameters default settings
Sterilization temperature: 134
º
C (273
º
F)
Sterilization time: 3.5 minutes.
Drying time: 1 minute.
Operations Sequence:
Pulse low/Pulse high:
Four vacuum pulses will build vacuum down
to 25 kPa.
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization:
Temperature and pressure are maintained constant at
the pre-set level 134ºC (273
º
F) for sterilization time (3.5 minutes).
Cooling:
N/A.
Fast Exhaust:
Steam is exhausted out of the chamber at a fast rate
until pressure decreases to ambient pressure.
Drying:
Vacuum is built up and the air pump is working.