Page 36
5.8 Program 8: Hollow Load (PV – pre-vacuum models
only)
For class B sterilization of pipes and wrapped pipes, double wraps,
when the manufacturer recommends autoclaving at temperatures of
121
C (250
F).
Nominal parameters default settings
Sterilization temperature: 121
C (250
F)
Sterilization time: 15 minutes
Drying time: 15 minutes.
Operations sequence:
Pulse low/Pulse high:
four vacuum pulses remove air by drawing
the chamber down to 25 kpa.
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
Cooling:
N/A
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Drying:
vacuum is built up and the air pump is working.