Corrections 

 

 

 

5.1 

Replace the DIG-T2 board. 

 

 

 

 

 

5.2 

Fix the connections. 

 

 

5.3 

Disassemble the valve and check that the 

plunger is free to move and that the seat 

is clean and free from debris. 

 

 

 

5.4 

Replace the ANL-T2 board

Possible cause check-up and tests 

The problem may be a mechanical failure of the 

valve, the valve driver located

 

on the  ANL-T2 

board or the control circuit on the DIG-T2 board. 

5.1 

Check TP9. If the voltage is higher than 

4V and the valve is open, then the DIG-

T2 board is faulty. If the voltage at TP9 

is <1VDC and the valve is open, the 

problem is not the DIG-T2, but some 

other element of the system.  

5.2 

Check the connection between the 

connector JP10 - 5 and the valve for a 

ground short.  

5.3 

Remove the cover protecting the 

terminals of the coil and check the 

voltage across the coil when the solenoid 

is connected. If the coil voltage is less 

than 2V and the valve is open, there is a 

mechanical problem

5.4 

If the voltage across the coil is higher 

than 10VDC, then the ANL-T2 board is 

faulty. 

 

     

10.    TROUBLESHOOTING 

(cont.)

 

      

  

10.4

    

In Process Malfunction

 (cont.) 

 

 

Symptom

 

5. 

Water inlet valve (21) to 

mineral free reservoir is 

always open  

  

  

Содержание 2540 EHS

Страница 1: ...TECHNICIAN MANUAL Electronic Table top Pre and Post Vacuum Autoclaves models 2540 3870 EHS Cat No MAN205 0063 005E Rev D...

Страница 2: ...ard 13 4 3 Digital Board DIG T2 13 4 4 Analog Board ANL T2 15 4 5 AC Board AC T1 17 5 CALIBRATION OF PRESSURE AND TEMPERATURE 19 5 1 Calibration Overview 19 5 2 Calibration Theory Error Compensation 2...

Страница 3: ...e Switch 60 9 13 The Heaters 61 9 14 The Dual Compartment Water Reservoir 62 9 15 The Water Sensing Electrodes 64 9 16 The Vacuum Pump 66 9 17 Draining the Jacket 68 9 18 Replacement of the Door Cover...

Страница 4: ...CTION OF THE 3870EHS AUTOCLAVES ELECTRONICS BOX 118 VESSEL ASSEMBLY 118 VESSEL ASSEMBLY 119 OUTER CABINET ASSEMBLY 120 TRAY HOLDER 121 TRAY 122 TRAY HANDLE CMT240 0001 122 DOOR TIGHTENING BOLT ASSEMBL...

Страница 5: ...he use of the technician It is strongly recommended that only qualified and Tuttnauer factory trained personnel service this autoclave and do so in accordance with the instructions in this manual Any...

Страница 6: ...with shut off valve and NPT male end Higher rates of pressure will require the installation of a pressure reducer A regulator and pressure gauge are supplied with the unit The minimum rate of flow is...

Страница 7: ...reducer A regulator and pressure gauge are supplied with the unit The minimum rate of flow is 0 26 gal min 1 lit min Installation of the valve should be 2 above counter height at the rear of the unit...

Страница 8: ...or requirement i e only non hazardous liquids shall be disposed of in public sewage Network and connection should comply with the devices consumption local installation and safety rules and regulation...

Страница 9: ...the wall to allow ventilation It is recommended that enough space be left around the autoclave to give a technician access for servicing the machine 3 6 Connections to Utility Supplies Using the hose...

Страница 10: ...9 REAR VIEW...

Страница 11: ...y water pressure into the unit is between 15 58 psi Verify that the mineral free water pressure into the unit is between 7 30 psi Door check ensure that the door locking mechanism is functioning prope...

Страница 12: ...without sediment Hardness 0 1 mmol l Attention The use of water in the autoclave that does not comply with the table above may have severe impact on the working life of the sterilizer and can invalida...

Страница 13: ...ower supply In addition the system is capable of direct communication with a printer over a parallel interface and or a PC via an RS 232 port for more detail see individual descriptions below CONTROL...

Страница 14: ...e ANL T2 analog board via ribbon cables The DIG T2 board receives human commands through the Keypad and transmits visual information back through the Keypad display The DIG T2 board receives sensory i...

Страница 15: ...e is running It forces the micro controller to recheck system inputs every 1 5 seconds this prevents the software from becoming stuck in a program loop that would lead to the software crashing and a l...

Страница 16: ...8 is connected to the Power Supply and receives the 12VDC and the 5VDC JP6 provides a connection to the DIG T2 board JP7 provides a connection for the Test Point board needed in troubleshooting JP9 pr...

Страница 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...

Страница 18: ...pply Three 230 115 VAC triac control circuits These are the AC drivers used in controlling the water and vacuum pumps and the steam valve The maximum amperage this board is capable of controlling is 6...

Страница 19: ...18 AC T1 BOARD Top View Bottom View...

Страница 20: ...ar output and has an electrical output range of 100mv 2400mv which corresponds to a temperature range of 68 F 20 C 302 F 150 C The temperature and pressure circuits provide analog input voltages that...

Страница 21: ...either F or C depending on what units the machine is set to display Note The pressure calculations are always done in Absolute pressure Absolute pressure atmospheric pressure 14 5 The calibration step...

Страница 22: ...slight delay in updating the screen when changing values on the PT100 simulator This will be the R value for the upper row Repeat this procedure for the high setting using the PT100 simulator it is s...

Страница 23: ...at the high end of the pressure range Once in the calibration mode the calibration screen will look similar to this Upper Row A 15 R 15 atmospheric pressure reading Lower Row A 45 R 45 high pressure...

Страница 24: ...ered the sterilization mode Record the temperature and pressure from the digital display Abort the cycle by pressing the STOP key Consult the Saturated Steam Table for the Absolute pressure that corre...

Страница 25: ...ave already added 14 5 Press the PROGRAM key one final time If PROGRAM WAIT has not been displayed then continue to press the PROGRAM key until it appears After a brief period the unit will return to...

Страница 26: ...nue pressing the program key until CALIB CODE 1 is displayed Using the UP arrow key select CALIB CODE 107 Press the PROGRAM key again until Program Wait is displayed this is an intermediate calibratio...

Страница 27: ...O RESERVOIR 21 0V Off 5V On TP10 OUTPUT WATER TO VAC PUMP 15 0V Off 5V On TP11 TP12 OUTPUT STEAM TO CHAMBER 93 0V CLOSE 5V OPEN TP13 OUTPUT SLOW EXHAUST 74 0V CLOSE 5V OPEN TP14 OUTPUT DRAIN COOLING W...

Страница 28: ...t ULKA water PUMP operates Air On Air valve 43 Air valve is open Water Res On Mineral free water reservoir valve 21 Water valve to mineral free reservoir is open Exh Res On Vacuum exhaust valve 52 Vac...

Страница 29: ...tes Float switch does not sense water and opens valve 21 6 3 Resetting the Autoclave Whenever it becomes necessary to restore the system to normal operation the system must be reset This will remove c...

Страница 30: ...29 6 4 Input Output Graph...

Страница 31: ...30...

Страница 32: ...31...

Страница 33: ...32...

Страница 34: ...33...

Страница 35: ...34...

Страница 36: ...83 49 8 22 233 07 39 53 284 70 9 23 235 49 40 54 285 90 10 24 237 82 41 55 287 07 11 25 240 07 42 56 288 23 12 26 242 25 43 57 289 37 13 27 244 36 44 58 290 50 14 28 246 41 45 59 291 61 15 29 248 40 4...

Страница 37: ...listed below proceed as follows a Select the program cycle to be modified by pressing the SEL CYCLE key b Press the PROGRAM key until CODE appears c Using the UP DN keys enter the access code for the...

Страница 38: ...Default Values Value 3870 3 5 8 30 8 3 5 8 2 3 Dry Time Time required for drying This parameter will set the time desired for drying Access Code 1 Resolution 1 minute Minimum Value minutes Maximum Val...

Страница 39: ...age The unit will signal a Cycle End only when the exhaust timer has reached zero and the unit is below the End Temp This also indicates the proper time to open the door and unload the chamber Access...

Страница 40: ...eed to achieve in the prevacuum stage This parameter is expressed in psia atmospheric pressure at sea level is 14 7psia A psia lower than 14 7 indicates a vacuum During a cycle the autoclave displays...

Страница 41: ...e 3 seconds Maximum Value 1800 seconds 30 minutes Cycle 1 2 3 4 5 6 Default Values Value 3 3 3 3 3 3 8 2 12 WtrResTime Timed extension for pumping mineral free water This parameter sets the length of...

Страница 42: ...F 273 F 5 4 F If limitP add is 2 the maximum allowable pressure will be 36 If the pressure reaches the upper limit as defined by limitP add i e 36 in our example the heating element will be switched...

Страница 43: ...nit based upon the SteTemp parameter For example if the required temperature is 250 F the required pressure calculated by the program is 16 psig This pressure can be increased by changing the value of...

Страница 44: ...his process will begin 4 minutes after the start of the Dry Stage and continue until the Dry stage is completed If Dry Vac 6 the Air Inlet Valve will be used to regulate the pressure during the Dry St...

Страница 45: ...and for every 100m below sea level the atmospheric pressure increases 0 14 psia Changes in pressure do to weather will not affect the accuracy of this unit Access Code 13 Resolution 0 1 psi Minimum Va...

Страница 46: ...Cycle 1 2 3 4 5 6 Default Values Value 0 7 0 7 0 7 0 7 0 7 0 4 8 2 20 PresInPsig selects the pressure display units This parameter sets the units of pressure that are printed by the printer and used o...

Страница 47: ...exhaust will be opened again for 1 second This process will continue until the temperature and pressure are correctly aligned Access Code 13 Resolution 1 Value 0 or 1 If Flash 0 This feature is not ac...

Страница 48: ...of the testing instrument should be 500V The insulation resistance should be at least 2 M The test is successful if there was no leakage 2 Protective Earth Impedance Test The test should be performed...

Страница 49: ...before removing outer covers 1 Remove the screws holding the rear cover 1 2 Remove the screws holding the cover to the base 2 3 Dismantle the air filter from the service opening cover 3 3 1 Remove th...

Страница 50: ...n the chamber or jacket 1 Take off the autoclave cover see para 9 2 Removing the Autoclave s Outer Covers 2 Remove the water reservoir gasket 3 Unscrew the safety valve and remove it from the safety v...

Страница 51: ...tch 7 8 Remove the DIG T2 and keypad from the plastic panel 9 Disconnect the DIG T2 from the keypad 10 Install a new DIG T2 to the keypad and reinstall it into the plastic housing 11 Follow the previo...

Страница 52: ...51 Electronic box 3870 Electronic box 2540 Control panel 7 2 6 3 2 4 5 7 1 1 Electronic box Electronic box 1 1 7 2 5 6 4 2 3 Control panel 2 7...

Страница 53: ...emoving outer covers 9 5 1 Replacing the O rings 1 Open the manual drain cock on the front of the machine and drain any water left in the reservoir 2 Continue turning the drain cock in a counterclockw...

Страница 54: ...er Covers of the Autoclave 2 Disconnect the drain tube from the valve using a 9 16 wrench 3 Remove the nut 3 and the ring for drain valve 2 4 Remove the drain valve 1 from the panel 5 Install a new va...

Страница 55: ...s not protrude from the doors inner surface 4 Operate the autoclave and verify that there are no leaks 5 Reinstall the door cover 9 6 2 Model 3870 The gauge is located in the plastic housing adjacent...

Страница 56: ...t 3 2 Gently pull out the door safety device locking pin 13 And using a bellows extraction tool remove the bellows assembly 4 It is possible that the washers 7 10 will be stuck if so again use compres...

Страница 57: ...nnect the flat cable 1 connecting the printer to the DIG T2 board 4 Unscrew the two screws 2 fastening the printer to the panel 5 Remove the printer and install the new one 6 Verify that the printer i...

Страница 58: ...place it with a new one 4 Reconnect the black ground wire 2 to the microswitch 5 Test the connection with an ohmmeter Connect the ohmmeter to the common terminal of the microswitch 1 and chassis groun...

Страница 59: ...he panel 4 Replace the circuit breaker with a new one 5 Reconnect the electrical wires 6 Reassemble the cover 7 Turn on the autoclave and verify it operates correctly 8 Move the circuit breaker s leve...

Страница 60: ...e para 9 2 Removing the Autoclave s Outer Covers 2 Disconnect the wires from the pump 3 Empty the water reservoir by using the drain value on the front of the autoclave 4 Disconnect the piping from th...

Страница 61: ...g so will void the warranty To replace the pressure switch perform the following steps Caution Before starting disconnect the instrument from the power source and ensure that there is no pressure in t...

Страница 62: ...e water from the jacket by gradually opening the drain valve at the back of the unit Wait until all the water has drained from the jacket see sec 9 17 3 Disconnect the electrical leads for each heater...

Страница 63: ...ervoir and a second float below it ensures that the water pump does not run if the compartment is empty The second compartment supplies feed water to the vacuum pump Water enters at the top through a...

Страница 64: ...astic tubing going to the reservoir 5 Disconnect any solenoid valves going to the reservoir 6 Drill out the pop rivets holding the legs of the reservoir to the chassis 7 Remove the reservoir from the...

Страница 65: ...the water drain valve located at the rear of the autoclave See sec 9 17 3 Remove the electrical connections from the terminals of the electrodes Mark the wires for each electrode to avoid miss wiring...

Страница 66: ...65...

Страница 67: ...age 1 Take off the autoclave cover see para 9 2 Removing the Outer Covers of the Autoclave 2 Close the external Feed Water Supply valve going to the back of the autoclave Count the number of turns and...

Страница 68: ...67...

Страница 69: ...hat the electric cord is disconnected and there is no pressure in the chamber or jacket 1 Turn off the main switch 2 Remove the rear cover 3 Turn the valve handle 1 1 4 turn and the jacket will begin...

Страница 70: ...as follows 4 1 Press the microswitch and listen to hear it click The clicking sound indicates that the microswitch has been activated 4 2 Close the door until the closing device feels tight also verif...

Страница 71: ...e the two Teflon disks 10 resting on the top and bottom of the lock housing axe 5 4 Position the new closing device on the lock housing axe 5 making sure that the Teflon disks 10 are in between the lo...

Страница 72: ...g the Outer Covers of the Autoclave 2 Unscrew the nut 3 attaching the sensor 2 to the fitting 1 3 Remove the sensor from the fitting 4 Disconnect the sensor s wire 4 from the electronic box and remove...

Страница 73: ...or by any other reason Before any operation or test on the autoclave perform a visual check for loose parts or for signs of over heating and burnt components 10 1 1 Manual Valves There are four 4 man...

Страница 74: ...age from the chamber or from the jacket 1 1 2 Faulty heating element 1 1 3 Faulty pressure switch 1 1 4 On model 2540 only 1 1 4 1 Cut off thermostat disconnects the heating element 1 1 5 On model 387...

Страница 75: ...Steam trap by pass needle valve is too closed 1 2 4 Slow exhaust valve 74 does not open 1 3 Temperature and or pressure are not calibrated 2 1 Pressure is high and temperature is low 2 1 1 Slow exhaus...

Страница 76: ...ber or from the jacket 3 1 2 Faulty heating element 3 1 3 Pressure switch faulty or adjusted to low pressure 3 1 4 On model 2540 only 3 1 4 1 Cut off thermostat disconnects the heating element 3 1 5 O...

Страница 77: ...ssure are not calibrated 3 2 3 Air inlet valve 43 is stuck open 3 2 4 Vacuum valve 52 is stuck open 3 2 5 Fast exhaust valve 73 is stuck open 3 2 6 Slow exhaust valve 74 is stuck open 3 2 7 Steam valv...

Страница 78: ...elements 5 3 Faulty pressure switch 5 4 Faulty cut off thermostat on model 3870 5 5 Lower electrode does not sense water 5 5 1 Faulty electrode 5 5 2 Faulty water pump 5 5 3 Water inlet pipe to the ja...

Страница 79: ...uum pump 7 3 Vacuum piping is leaking 7 4 Non return valve is stuck 7 5 Vacuum valve 52 does not open during vacuum stage due to 7 5 1 Faulty coil or plunger 7 5 2 Pump disconnected 7 5 3 Faulty digit...

Страница 80: ...voltage must be 208 VAC With a multimeter check from pin 3 at the AC INPUT connector JP3 on the AC T1 board to pins 1 and 2 Turn on the power The reading must be the same as the voltage coming from th...

Страница 81: ...nector on the ANL T2 board Connect the negative probe of a multimeter to TP1 on the test board With the positive probe check TP2 for 5V DC an acceptable range is 4 9 5 3V With the positive probe check...

Страница 82: ...hen proceed down the line to the device attached to the other end of the cable or connector and continue with the previous step The problem could be with any one of the external devices connected to t...

Страница 83: ...s are OK the display is probably damaged 3 1 DIG T2 board is faulty 4 1 Keyboard is faulty 4 2 DIG T2 board is faulty 5 1 Check if the displayed value is near the atmospheric pressure 5 2 If after set...

Страница 84: ...ading of the reservoir float switch position 7 1 Check that the electrodes are sensing water in the jacket 10 TROUBLESHOOTING cont 10 3 Pre process malfunction cont Symptom 6 Pressing the START key do...

Страница 85: ...re in the chamber rises during the heat up stage it presses outward on the door This can cause the door switch to be released and cause the system to think the door is opened 2 1 Fast exhaust valve 73...

Страница 86: ...DIG T2 board is faulty If the voltage at TP8 is 1VDC and the valve is open the problem is not the DIG T2 but some other element of the system 3 2 Check the connection between the connector JP10 4 and...

Страница 87: ...DIG T2 board 4 2 Check the connection between connector JP10 4 and the valve for an open circuit 4 3 Check the coil input voltage with the solenoid connected to the valve If the voltage is higher tha...

Страница 88: ...he DIG T2 board is faulty If the voltage at TP9 is 1VDC and the valve is open the problem is not the DIG T2 but some other element of the system 5 2 Check the connection between the connector JP10 5 a...

Страница 89: ...nnection between connector JP10 5 and the valve for an open circuit 6 3 Check the coil input voltage with the solenoid connected to the valve If the voltage is higher than 10VDC and the valve does not...

Страница 90: ...be open then the problem is on DIG T2 board 7 2 Check the connection between connector JP10 3 and the valve 7 3 Check the coil input voltage with the solenoid connected to the valve If the voltage is...

Страница 91: ...ve is open then the DIG T2 board is faulty If the voltage at TP7 is 1VDC and the valve is open the problem is not the DIG T2 but some other element of the system 8 2 Check the connection between the c...

Страница 92: ...TP4 if the voltage is higher than 4VDC the problem is on DIG T2 board 9 2 If the voltage at TP4 is 1VDC the problem could be either the ANL T2 or the SSR Turn power off and remove the wire from termin...

Страница 93: ...voltage at TP4 is always 1VDC then check if the electrodes are sensing water in the jacket If they are then the problem is the DIG T2 board If the voltage at TP4 is higher than 4VDC the problem could...

Страница 94: ...k the fuse at the rear of the unit 12 4 With the power off check if the pump is free to rotate Do this by inserting a flat blade screwdriver into the center of the rear bell housing engaging a slotted...

Страница 95: ...and check its resistance At 32 F the resistance across the two wires of the sensor should read 100 The resistance at higher temperature can be calculated by adding 0 2 ohms for each degree F 13 4 Plug...

Страница 96: ...he sensor is properly plugged into JP2 on the ANL T2 board 15 2 Check the positioning of the PT100 see sec 9 20 If the positioning is correct then the PT100 sensor is faulty 15 3 The difference in the...

Страница 97: ...the ANL T2 board 17 2 Check that the wires and connector are in good condition 17 3 The pressure sensor is defective 17 4 The pressure measuring circuit on ANL T2 board is defective 18 1 Check that th...

Страница 98: ...Replace the DIG T2 Possible cause check up and tests 20 1 Check if the float moves up and down freely 20 2 Check that the wires and connectors at the float and the connector at JP11 and JP12 on the AN...

Страница 99: ...approx 255 Short circuit the electrode by touching the electrode housing with the wire of the disconnected electrode The displayed number should be 0 If so the electrode is faulty 21 3 Check for an o...

Страница 100: ...If the voltage at TP10 is 1VDC and the valve is open the problem is not the DIG T2 but some other element of the system 24 2 Check the connection between the connector JP10 6 and the valve for a grou...

Страница 101: ...problem is on DIG T2 board 25 2 Check the connection between connector JP10 6 and the valve for an open circuit 25 3 Check the coil input voltage with the solenoid connected to the valve If the volta...

Страница 102: ...is constantly 1VDC then the problem is on the DIG T2 board 26 4 If the voltage at TP17 is higher than 4DC then remove the JP7 connector and check the output of JP7 2 on the AC T1 board to ground 26 5...

Страница 103: ...hen the problem is on DIG T2 board 28 2 Check the connection between connector JP10 8 and the valve for an open circuit 28 3 Check the coil input voltage with the solenoid connected to the valve If th...

Страница 104: ...lve is open then the DIG T2 board is faulty If the voltage at TP7 is 1VDC and the valve is open the problem is not the DIG T2 but some other element of the system 29 2 Check the connection between the...

Страница 105: ...is open then the DIG T2 board is faulty If the voltage at TP13 is 1VDC and the valve is open the problem is not the DIG T2 but some other element of the system 30 2 Check the connection between the co...

Страница 106: ...open then the problem is on DIG T2 board 31 2 Check the connection between connector JP10 9 and the valve 31 3 Check the coil input voltage with the solenoid connected to the valve If the voltage is...

Страница 107: ...the pump 32 3 Check the vacuum valve 52 is open allowing the pump to draw from the chamber 32 4 Check the vacuum pump With an independent gauge installed in the door port verify that the pump is pulli...

Страница 108: ...it between terminals 3 4 and 1 2 and 3 1 and 4 2 and 3 2 and 4 1 Make sure there are no direct shorts 34 2 Next using an ohmmeter check for an open between terminals 1 2 be sure to reverse the meter l...

Страница 109: ...th the device in question in the OFF mode pressing the PROGRAM KEY to display the date and time will turn the heaters off a voltage reading from terminal 1 to ground will read approximately 110 VAC A...

Страница 110: ...e ANL T2 board or the control circuit on the DIG T2 board 35 1 Check the connection between the connector JP10 12 and the valve for a ground short 35 2 Remove the cover protecting the terminals of the...

Страница 111: ...10 12 and the valve for an open circuit 36 2 Check the coil input voltage with the solenoid connected to the valve If the voltage is higher than 10VDC and the valve does not open then the solenoid or...

Страница 112: ...he seat of the valve has become dirty from a mineral build up 2 2 Safety relief valve is faulty 3 1 Safety relief valve is faulty 4 1 Safety relief valve is faulty 5 1 Safety relief valve is faulty 6...

Страница 113: ...s steam in the jacket by carefully pulling the ring of the safety valve 7 2 The heating elements are not producing enough power 8 1 Check for a vacuum leak use program 6 If maximum vacuum is reached b...

Страница 114: ...and tests Check pipes connecting the water reservoir filter and water pump 1 1 There is an air pocket at the pump inlet 1 2 Pipe is clogged 1 3 Filter is clogged See section 10 4 26 See section 10 4 2...

Страница 115: ...LE035 0055 ELE035 0055 14 Holder Safety Valve E M CMT100 0002 CMT100 0002 15 Air Jet M E Black CMT100 0006 CMT100 0006 16 Valve Safety Steam 1 4 40 psi ASME SVL029 0004 SVL029 0004 17 Water Level Elec...

Страница 116: ...ver RCV254 0009 RCV387 0002 44 Cover Printer Opening Superp POL067 0002 POL067 0002 45 Transducer Pressure MPX 2200 AP THE006 0003 THE006 0003 46 Power Supply PM65 23A ELE035 0032 ELE035 0032 47 Relay...

Страница 117: ...g Bolt Door Locking Bellows LOK240 0026 LOK240 0026 69 Housing Door Locking Bellows LOK240 0025 LOK240 0025 70 Membrane pin L0K240 0023 L0K240 0023 71 Inner bushing for bellow CMT067 0002 CMT067 0002...

Страница 118: ...Digital Input Electrodes Chamber temperature sensor Vacuum Pump AC Control Water Pump AC Control PC Communication Port Test Points Connector Jacket Pressure sensor Chamber Pressure sensor Pressure sw...

Страница 119: ...Input Electrodes Chamber temperature sensor PC Communication Port Test Points Connector Jacket Pressure sensor Chamber Pressure sensor Pressure switch Low Float of Vacuum Pump Reservoir Cut Off Switch...

Страница 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...

Страница 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...

Страница 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...

Страница 123: ...122 TRAY TRAY HANDLE CMT240 0001 MODEL SIZE CAT No 2540 TRY240 0001 Big TRY387 0001 3870 Small TRY387 0003...

Страница 124: ...OK240 0012 LOK387 0012 5 Locking housing axe LOK240 0014 LOK387 0014 6 Door locking device pin LOK240 0019 LOK387 0016 7 Bakelite handle HAN071 0003 HAN071 0006 8 Closing bridge c clip NUT193 0339 NUT...

Страница 125: ...manual Size mm 2540 3870 1 FEED WATER 13 Feed water cool fast exhaust 14 Feed water to reservoir 15 Feed water to vacuum pump 3 2 4 4 5 3 4 5 4 5 2 MINERAL FREE WATER 21 Mineral free water inlet 3 4 5...

Страница 126: ...125 PIPING DRAWING FOR 2540 EHS...

Страница 127: ...126 PIPING DRAWING FOR 3870 EHS...

Страница 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...

Страница 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...

Страница 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...

Отзывы: