47
9
MAINTENANCE AND REPLACEMENT PROCEDURES
9.1
Safety Tests after Repair
ATTENTION!
After every repair or dismantling of the enclosure, it is
recommended that the following tests be performed:
1. Enclosure Current Leakage Test.
The test should be performed using a Megger. Make sure the
autoclave is unplugged. The electrical potential of the testing
instrument should be 500V. The insulation resistance should be at
least 2 M
Ω
.
The test is successful if there was no leakage.
2. Protective Earth Impedance Test
The test should be performed using an Ohmmeter. Make sure the
autoclave is unplugged from the wall outlet. Next measure the
resistance between the grounding screw on the rear plate (or any
other unpainted metallic part) and the grounding pin of the power
cord plug. The resistance should not exceed 0.3
Ω
, a high
resistance would indicate a faulty ground connection.
Содержание 2540 EHS
Страница 10: ...9 REAR VIEW...
Страница 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Страница 19: ...18 AC T1 BOARD Top View Bottom View...
Страница 30: ...29 6 4 Input Output Graph...
Страница 31: ...30...
Страница 32: ...31...
Страница 33: ...32...
Страница 34: ...33...
Страница 35: ...34...
Страница 66: ...65...
Страница 68: ...67...
Страница 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Страница 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Страница 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Страница 123: ...122 TRAY TRAY HANDLE CMT240 0001 MODEL SIZE CAT No 2540 TRY240 0001 Big TRY387 0001 3870 Small TRY387 0003...
Страница 126: ...125 PIPING DRAWING FOR 2540 EHS...
Страница 127: ...126 PIPING DRAWING FOR 3870 EHS...
Страница 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Страница 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Страница 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...