24
— Press
the
PROGRAM
key to advance to the next
value.
— The
cursor
on
the
bottom
row
will
move one
position to “R”;
now
enter the pressure value
recorded from the digital display to which you
have already added 14.5.
— Press
the
PROGRAM
key one final time
If PROGRAM WAIT has not been displayed then
continue to press the
PROGRAM key
until it appears.
After a brief period the unit will return to a Ready
state
This completes the calibration procedure for
temperature e and pressure. A sterilization cycle
should be performed to verify that the system is
calibrated and running properly.
Содержание 2540 EHS
Страница 10: ...9 REAR VIEW...
Страница 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Страница 19: ...18 AC T1 BOARD Top View Bottom View...
Страница 30: ...29 6 4 Input Output Graph...
Страница 31: ...30...
Страница 32: ...31...
Страница 33: ...32...
Страница 34: ...33...
Страница 35: ...34...
Страница 66: ...65...
Страница 68: ...67...
Страница 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Страница 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Страница 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Страница 123: ...122 TRAY TRAY HANDLE CMT240 0001 MODEL SIZE CAT No 2540 TRY240 0001 Big TRY387 0001 3870 Small TRY387 0003...
Страница 126: ...125 PIPING DRAWING FOR 2540 EHS...
Страница 127: ...126 PIPING DRAWING FOR 3870 EHS...
Страница 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Страница 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Страница 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...