11
3.8
Water Quality
3.8.1
Built-in Steam Generator
The distilled or mineral – free water supplied to the steam
generator shall be according to the table below:
A Reverse Osmosis system meeting the qualifications below
may be used to provide water for the steam generator. The
better the quality of the water, the better performance, the less
maintenance and the longer the life of the autoclave.
Mineral Free Water qualifications
(In compliance with ISO 11134 and ISO 13683)
Evaporate residue
≤
15 mg/l
Silica
≤
2 mg/l
Iron
≤
0.2mg/l
Cadmium
≤
0.005 mg/l
Lead
≤
0.05 mg/l
Rest of heavy metals
≤
0.1 mg/l
Chloride
≤
3 mg/l
Phosphate
≤
0.5 mg/l
Conductivity
≤
50 µs/cm
pH
6.5 to 8
Appearance
Colorless, clean, without sediment
Hardness
≤
0.1 mmol/l
Attention:
The use of water in the autoclave that does not comply with
the table above may have severe impact on the working life
of the sterilizer and can invalidate the manufacturer’s
warranty.
The suitability of the mineral free water to be used should be
verified by testing in accordance with the above table; at an
authorized laboratory using acknowledged analytical
methods. We recommend testing the water quality once a
month.
3.8.2
Water for the Vacuum System and the Drain Cooling
The feed water supplied, typically from a public water system,
to the liquid ring vacuum pump must meet the following
requirements:
♦
Hardness: 0.7 - 2 mmol/l.
♦
Water temperature: shall
not exceed 59°F (15°C).
Note: The use of hard water in the vacuum pump may invalidate
the warranty for the vacuum pump, since it can cause
blocking of the rotor, which can damage the pump.
Содержание 2540 EHS
Страница 10: ...9 REAR VIEW...
Страница 17: ...16 Layout diagram of the ANL T2 is provided below ANL T2 BOARD Top View Bottom View...
Страница 19: ...18 AC T1 BOARD Top View Bottom View...
Страница 30: ...29 6 4 Input Output Graph...
Страница 31: ...30...
Страница 32: ...31...
Страница 33: ...32...
Страница 34: ...33...
Страница 35: ...34...
Страница 66: ...65...
Страница 68: ...67...
Страница 120: ...119 VESSEL ASSEMBLY Model 2540 EHS Model 3870 EHS...
Страница 121: ...120 OUTER CABINET ASSEMBLY Model 2540 EHS Model 3870 EHS...
Страница 122: ...121 TRAY HOLDER For model 2540 TRH254 0009 For model 3870 TRH387 0001...
Страница 123: ...122 TRAY TRAY HANDLE CMT240 0001 MODEL SIZE CAT No 2540 TRY240 0001 Big TRY387 0001 3870 Small TRY387 0003...
Страница 126: ...125 PIPING DRAWING FOR 2540 EHS...
Страница 127: ...126 PIPING DRAWING FOR 3870 EHS...
Страница 128: ...127 ELECTRICAL DRAWING FOR 2540 EHS...
Страница 129: ...128 ELECTRICAL DRAWING FOR 3870 EHS TILL S N 2307039...
Страница 130: ...129 ELECTRICAL DRAWING FOR 3870 EHS AFTER S N 2307040...