Chapter 3
44
-HUMIDITY CAL: STEP 2-
SET TO CORRECT
HUMIDITY VALUE
35.1
% RH
v
↵
TO ACCEPT
AND CONTINUE
Figure 4–17: Humidity Calibration Screen
Note
: If the two readings are the same, there is no need to proceed
with the calibration. Press the
&
key to terminate the
calibration procedure.
5. Q-T
RAK
Plus monitor will check to see if the measured parameter is
stable. The screen displays the parameter in real time to show any
fluctuations.
6. At the successful conclusion of the calibration, the screen will highlight
"CAL COMPLETE". Press the
↵
key to continue.
If an error is detected during the calibration process, an error message will
be displayed. The previous calibration values are kept. Press the
&
or the
↵
key to return to the User Calibration Menu.
Auxiliary Sensor Calibration
To perform a calibration on the auxiliary sensor on the Q-T
RAK
Plus
monitor, please refer to the documentation sent with the auxiliary sensor.
Restore Factory Calibration
On very rare instances, it may be necessary to restore the calibration of the
instrument to the original factory settings. This should only be used when
absolutely necessary.
1. To restore the factory calibration, select
RESTORE DEFAULTS
in the
Calibration Select page by using the
!
or
"
key and press the
↵
key.
2. The following warning message appears on the display. Press the
"
key
to continue or the
↵
key to abort (Figure 4-18).