Chapter 3
40
with a known concentration of CO for the span calibration, 35 ppm or
200 ppm is recommended. Contact your local TSI distributor for available
CO calibration kits (Figure 4-11).
Notes
:
!
Calibration with CO gas should always be done in a well-
ventilated area.
!
100% Nitrogen cannot be used as a zero gas for the CO
sensor.
The calibration procedure can be aborted at any time by pressing the
&
key.
1. To put the Q-T
RAK
Plus monitor into CO calibration mode, press the
!
or
"
keys to select
CALIBRATION
in the Setup Menu.
2. Select CO by pressing the
!
or
"
keys in the calibration submenu and
press the
↵
key.
Locate the calibration collar and slide it over the main probe (Refer to
Figure 4–11).
Figure 4–11: CO Calibration