Chapter 3
20
To program the Q-T
RAK
Plus monitor for Log Mode 2 or 3,
enter the following:
Item Description
Start Date
Start Time
Enter the date and time to begin the sample:
•
If the start date field is left blank, the date defaults to
the current date. The test starts at the start time.
•
If both of the start date and time fields are left blank,
the test begins as soon as you press the
↵
key in the
appropriate log mode.
Log Interval
Enter the log interval to use for the test.
Test Length
Enter the length for the sample:
•
The instrument automatically turns off when the last
test is complete at the end of the test length.
•
If the field is left blank, you must manually stop the
sample by pressing the
↵
key in the logging screen.
Number of
Tests
Enter the number of tests to perform.
Time between
tests
Enter the time between tests, if there is more than one
test. If 0 or blank is entered, the next test is started
immediately after the previous test.
While entering values for Log Mode 2 or 3, the “Percent
Memory Required” is dynamically updated to show the
amount of logger memory required to take the programmed
sample. If the protocol defined requires more than 100% of
memory, decrease the amount of memory required by
changing one of the following parameters:
•
Increase the logging interval.
•
Decrease the length for the test.
•
Decrease the number of tests.
The settings for each separate Log mode must
not
require
more than 100% of the logger memory.
Note:
If the percent memory required is greater than
the available memory, the TrakPro software will
not allow you to program Log Mode 2 or 3.
3.
When finished defining parameters for LOG 2 and LOG 3
modes, select
Send
.
4.
Disconnect the Q-T
RAK
Plus
monitor and cable from the
computer.
5.
The instrument is ready for data logging.
Содержание Q-TRAK Plus 8552
Страница 2: ......
Страница 16: ...Chapter 2 10 ...
Страница 40: ...Chapter 3 34 ...
Страница 52: ...Chapter 3 46 ...
Страница 57: ...TSI Incorporated ...