Chapter 3
22
To program this logging example, do the following:
1. Make sure the Q-T
RAK
Plus monitor is connected to the
computer and turned on.
2. Select
Logging Setup
from the
Instrument Setup
menu.
T
RAK
P
RO
software retrieves the current settings for Log
Modes 2 and 3 from the Q-T
RAK
Plus monitor and displays
them (as shown in the previous dialog box).
3. Enter the following for Log Mode 2 and 3:
Setting
LOG 2
LOG 3
Start Date
today's date
today's date
Start Time
five minutes
from now
five minutes
from now
Log interval
01:00
05:00
Test length
00:08:00 00:08:00
Number of tests
1
4
Time between tests 00:00:00
00:16:00
4. Select
Send
. The logging instrument is programmed for the mode
2 and mode 3 protocols.
5. Note in the example that the LOG 2 test requires 4% of the
available memory and LOG 3 requires 4% of the memory. A total
of 100% of the memory is available for use.
6. Disconnect the Q-T
RAK
Plus monitor from the computer. Refer to
other sections of this Operation and Service Manual for details on
making measurements using LOG 2 and LOG 3 modes.
After programming the Q-T
RAK
Plus monitor with the T
RAK
P
RO
TM
Data Analysis Software, take the monitor to the desired location and
turn it on. From the Main Menu, put the monitor into LOG MODE 2 or
LOG MODE 3 (whatever was programmed) by pressing the
#
or
$
key. Press the
↵
key to initiate the Log Mode. The display will show
the current time and date.
The following will happen, depending upon the protocol: