35
7200790 Service Manual for the HELION
®
XENION
®
Surgical Lighting System GB
11 Testing the Electrics of the HELION M/M+ / L/L+ Lamp
11.2
11.2
11.2
11.2
11.2 TTTTTesting the voltages and cables in the
esting the voltages and cables in the
esting the voltages and cables in the
esting the voltages and cables in the
esting the voltages and cables in the
lamps (cont.)
lamps (cont.)
lamps (cont.)
lamps (cont.)
lamps (cont.)
WARNING – Electric Shock:
The following measurements must be made
with the power supply switched on.
Cordon off installation site and proceed
with great care.
8. Switch on the power supply and the lamp and, if
necessary, set the lamp to maximum brightness.
Measure the voltage at the contacts of the filter glass
assembly:
9. Using a voltmeter, measure the voltage at the con-
tacts
B
:
• The voltage must be 23 V +/- 0.3 V.
-
In the case of voltage deviations, set the voltage
in accordance with Section 11.1.
If there is no voltage:
• Check the proper positioning of the cables:
• Check the cables for breaks:
-
Replace defective cables/components and fit as
described above, but in reverse order.
Installing the lamp cover:
10.Insert the sheet-metal clips
5
into the lamp cover
3
and fit the lamp cover
3
as described above,
but in reverse order.
11.Check the secure positioning of the lamp cover
3
.
Figure 03