7200790 Service Manual for the HELION
®
XENION
®
Surgical Lighting System GB
5 Changing the Lamp
Figure 03
5.3
5.3
5.3
5.3
5.3
Changing the lamp of the XENION M/
Changing the lamp of the XENION M/
Changing the lamp of the XENION M/
Changing the lamp of the XENION M/
Changing the lamp of the XENION M/
M+
M+
M+
M+
M+
Removal:
1. Unlock the central hand grip
2
by means of the
unlocking device
1
and remove it.
2. Undo the three screws
3
and remove the focus
unit
5
.
3. Remove the old metal vapour lamp
4
by pulling
it upwards and dispose of it in an appropriate man-
ner.
Installation:
4. Fit the new metal vapour lamp
4
holding it by
the packaging film/using a soft cloth and assemble
the components as described above, but in the
reverse order.
5. Insert the focus unit
5
into the lamp so that both
the plug connectors
6
are aligned.
6. Tighten the three screws
3
by hand.
7. Attach the central hand grip
2
.