7200790 Service Manual for the HELION
®
XENION
®
Surgical Lighting System GB
92
16 Technical Data (cont.)
Lighting data
HELION L
HELION L+
HELION XL+
Central luminous intensity at a defined distance
electrically dimmable from/to [%]
External dimmer
required
40 - 100
40 - 100
Central luminous intensity at a defined distance [lx]
50.000 - 140.000 50.000 - 140.000 60.000 - 145.000
Focusable luminous field size d10 at distance of 1 m [mm]
160 - 250
160 - 250
200 - 350
Focusable luminous field size d50 at distance of 1 m [mm]
83 - 165
83 - 165
104 - 175
Colour temperature [K]
4.300
4.300
4.300
Colour rendering index [Ra (1-8)]
93
93
93
Red rendering index [R9]
ca. 60
ca. 60
ca. 60
Total radiation intensity at 100,000 Lux [W/m²]
355
355
355
Luminous efficacy of radiation [lm/m²]
282
282
282
Illumination depth without refocusing [L1 / L2]
430 / 750
430 / 750
340 / 900
Brightness of reserve light [%]
100
100
100
MODELLING: residual illumination with 1 shutter [%]
40
40
64
MODELLING: residual illumination with 2 shutters [%]
41
41
53
MODELLING: residual illumination with tube [%]
100
100
87
MODELLING: residual illumination with tube and 1 shutter [%]
40
40
50
MODELLING: residual illumination with tube and 2 shutters [%]
41
41
42
Leneman laminar flow index
28
28
93
Oostlander laminar flow index
32
32
119
Electrical data
HELION L
HELION L+
HELION XL+
Transformer [mm/L x W x H]
120 x 103 x 118
120 x 103 x 118
120 x 135 x 118
Rated output of transformer [VA]
220
220
410
Supply voltage at transformer [V]
115/120/230/240 115/120/230/240 115/120/230/240
Maximum current consumption at 24 V DC [A]
6,6
7,5
13,5
Maximum current consumption at 230 V [A]
0,9
1,2
1,6
Maximum transformer permanent output [W]
185
195
375
Voltage at ceiling fixture point [V]
24-26 (AC/DC)
24-30 (DC)
24-30 (DC)
Ultimate voltage at lamp [V]
23
23
23
Rated output of lamp [W]
150
150
250
Rated voltage of lamp [V]
24
24
24
Effective service life of lamp [h]
600-1000
600 - 1000
600-1000
Power input of complete lamp [VA]
205
210
310
Voltage stability (corrected electronically)
yes
±0,1
±0,1
Lamp-friendly soft start
yes
yes
yes
Focus/luminous field memory
no
yes
yes
Automatic lamp changeover
yes
yes
yes
Electrics as per VDE and IEC
yes
yes
yes
Protection class as per MPG (Medical Devices Law)
I
I
I
Data for spare lamp
24V/150W
24V/150W
24V/250W
Spare lamp
8800100
[pack of 6 lamps]
8800100
[pack of 6 lamps]
8800250
[pack of 6 lamps]
Certification mark
CE · UL/CSA
CE · UL/CSA
CE · UL/CSA
Technical data
The technical data can be found in the data sheets in Section 15.1.