5–8
CUTOFF LIMIT AS A PERCENTAGE.
To enter the cutoff limit
as a percentage of the initial sample mass, press
〈
2
〉
at the
Cutoff
Units
menu shown on page 5–7. The oven displays:
Current Cutoff Limit
X.XXX %
do you want to
change Cutoff Limit?
f
To exit without changing the cutoff limit, press
〈
NO
〉
. The oven
returns to the
Burn Parameter Menu
.
f
To change the cutoff limit, press
〈
YES
〉
. The oven displays:
Enter Cutoff Limit
from 0.001% - 1.000%
in 0.001% steps
then press ENTER
Use the numeric and decimal keys to enter a cutoff limit from
0.001% to 1.000% in increments of 0.001%. Use the numeric
keys to set the burn time. After entering the cutoff limit, press
〈
ENTER
〉
. The oven returns to the
Burn Parameter Menu
.
Calculation Type
The oven provides two methods for calculating
%Loss
: the
bitumen-
to-sample
method and the
bitumen-to-aggregate
method. Both
methods measure the difference in the sample weights before and
after the burn. However, the
bitumen-to-sample
method expresses
the difference as a percentage of the
starting mass
while the
bitumen-to-aggregate
method expresses the difference as a
percentage of the
final (aggregate) mass
.
Содержание NTO 4730
Страница 10: ...x LIST OF TABLES Table Title Page 4 1 Oven Keys 4 3 5 1 Recommended Minimum Burn Times 5 6...
Страница 30: ...2 10 NOTES...
Страница 90: ...5 36 NOTES...
Страница 110: ...6 20 NOTES...
Страница 130: ...Appendix A 10 Figure A 1 Printer with Doors Open Figure A 2 Inserting the Paper Roll SPRING CLIPS...
Страница 136: ...Appendix A 16 NOTES...
Страница 148: ...Appendix C 6 NOTES...
Страница 156: ...Index 8 NOTES...
Страница 157: ...Troxler NTO Warranty 1 WARRANTY NOTES...