The oven verifies the scale accuracy, computes the error percentage,
then displays the results as shown:
Verification Result
XX.XXX %
do you want to
print result?
To print the results, press
〈
YES
〉
; to continue without printing, press
〈
NO
〉
or
〈
ESC
〉
.
If the error percentage is less than
±
0.1%, the oven returns to the
Scale Menu
. If the error percentage is greater than
±
0.1%, the
oven displays:
do you want to
calibrate?
f
To continue without calibrating the scale, press
〈
NO
〉
or
〈
ESC
〉
.
The oven returns to the
Scale Menu
.
f
To calibrate the scale, press
〈
YES
〉
. The oven proceeds as
described in the
Calibrate
section that begins on the following
page.