5
2.
Product Specification
Safety Classification
Electric
protection
Type
Class I
Electric
protection
Level
Defibrillation proof type CF applied Part
Ingress Protection
IP34 (Prevent solid objects larger than 2.5mm in diameter and water intrusion
from splashing in all directions)
Working mode
Continuous
Classification
Portable equipment, non-portable infusion pump
Specification Parameters
Infusion
line
specification
10-60 drops
System Accuracy
±
5%
Drip Infusion Rate
Accuracy
±
5% or
±
1drops/min
Infusion Rate
0.10-1500ml/h
Drip mode range
1~500drops/min
Bolus Rate
0.1-1500ml/h
Bolus preset
0.1-50ml
KVO Rate
0.1-5.00ml/h
Micro mode
setting range
100-1500ml/h
Minimum flow rate
increment
0.01ml/h
VTBI
0-9999.99ml, minimum step is 0.01ml
Total Volume
Infused
0-9999.99ml, minimum step is 0.01ml
Time Range
1min-99hrs59min
Fuse Type
slow fuse 2A 250V
Dimensions
199(W)*111(D)*126.5(H) mm
Weight
1.4kg
Power Supply
AC power supply
100-240V 50/60Hz
Input power
50VA
Содержание IP12
Страница 1: ...User Manual Triumph IP12 Infusion Pump ...
Страница 32: ...32 ...