23
Unit Mark
Unit Conversion
kPa
1 kPa=7.5mmHg=0.145psi=0.01bar
PSI
1psi=51.713mmHg=6.895kpa=0.069bar
Bar
1bar=787.5mmHg=15.225psi=105kPa
Bubbles Size
Click
『
Bubbles size
』
to enter into air bubble size setting interface, move the long box to the preset
level, confirm and then click
『
OK
』
.
The air bubble detector has 7 levels, when the volume of single air bubble or the total air bubbles
within 15min in the line reach the preset air bubble alarm threshold value, it will activate air
bubble alarm. The air bubble testing sensitivity is 20ul. It is suggested to select a suitable level
according to the actual requirement.
Air Bubble detector level
Alarm Threshold Value
Level 1
50ul
Level 2
100ul
Level 3
200ul
Level 4
300ul
Level 5
450ul
Level 6
600ul
Level 7
800ul
Reminder Alarm
Click
『
Reminder alarm
』
to enter into the time for reminder alarm setting interface, click the preset time
option to set the reminder alarm time.
Finish Pre-Alarm
Time for pre-alarm refers to the time of activating near completion alarm when the fluid/drug
infused volume is nearly reaching the preset value.
Click
『
Finish pre-alarm
』
to enter into the time for pre-alarm setting interface, click the preset time
option,to set the finish pre-alarm time.
Drop Sensor
Click
『
Drop sensor
』
to set ON or OFF.
The “Drop error “ alarm function is only available only when the drop sensor is installed (not included)