Infusion Mode Setting
After starting the equipment and self-test, the equipment automatically enters into the rate mode
parameters setting interface. To select the other modes, click
『
Menu
』
icon
to enter into the
main interface. Click
『
Modes
』
icon to enter into the mode selection menu interface, and select
preset infusion mode.
Rate Mode
Under this mode it allows you to set rate and volume to be infused. Set the two parameters, and
the system will automatically calculate the third parameter. If volume to be infused is “0” pump
will work until pumps alarms.
Time Mode
Under this mode, it allows to set VTBI (Volume to be infused) and time. The system will
automatically calculate the speed, speed = Volume(ml)
/
time(min)
Body Weight Mode
Under this mode, set the Weight (body weight), Active agent (drug mass), Conc. unit
(concentration unit), Volume (fluid volume), Dose rate, Dose unit, VTBI.
The system will automatically calculate the flow rate from the specified dose rate (ug/kg/min,
mg/kg/min, ug/kg/h, mg/kg/h,…etc) according to related formula {dose rate × weight}/{Active
agent (drug mass)/Volume(fluid volume)}, and automatically calculate the time according to
(VTBI) /(flow rate).