17
Changing the Rate During infusion
Under the running interface, click rate number in touch screen and reset target infusion rate. Or
【
Stop
】
,
and enter the parameters setting interface, then reset target infusion rate.
Note:
●
Only the
rate mode, drip mode
and
Body weight mode
support online rate
modification function during infusion without pressing
【
Stop
】
Bolus Application
In operation, Bolus functions have two operation modes: Manual bolus and Automatic bolus.
(1)
Manual bolus
: press and hold the
【
Bolus
】
button on product panel, pump will work at
the max flow rate or set max bolus rate under the setting interface, release the button, pump will
go back to the previous setting infusion rate.
(2)
Automatic bolus
: Under the running interface, click
『
Bolus
』
on touch screen, set two
parameters among bolus infusion volume, rate and time, click
『
Start
』
. It will beep at every 1ml
infused. After bolus infusion is finished, the equipment goes back to the previous infusion rate.
Infusion Completion
When remaining infusion time is near preset volume to be infused completion time, pump will
alarm. If ignored, the system will keep alarming until complete VTBI infusion.
After VTBI completed, it activates VTBI infused alarm, if KVO function is ON, the equipment
automatically starts KVO function, click
『
OK
』
in the alarm interface to stop KVO and eliminate
alarm.
The default working time of the KVO system is 30min, after reaching the time, it will activate
KVO completion alarm and stop infusion.