24
Note:
●
The default state for drop sensor function system is OFF. It can be manually turned
on by the user when the drop sensor should be adopted. If the function is ON when the drop sensor
is not installed, then the system will report “Drop sensor connection” alarm.
Drop sensor level
The sensitivity of the drop sensor can be adjusted in three levels. The higher the level is, the more
sensitive the detection will be.
Micro Mode
Click
『
Micro mode
』
to enter into micro mode setting interface. ON/OFF is optional in this function.
Under the ON mode, set the rate limit, then the infusion rate under any infusion mode is not
allowed to exceed this limit.
Reset Total Volume
Click
『
Reset total volume
』
, the interface displays the operation confirming prompt box, click
『
Yes
』
to confirm reset, otherwise, please click
『
No
』
General
In the main interface, click
『
General
』
to enter into the General equipment setting interface.
Date & Time
Click
『
Date &Time
』
to enter into the date and time setting interface. It allows to set the date, time
and format in this interface.
When setting date and time, directly input the numerical value in the input method interface. For
example, to preset one date “10-24-2019”, input “10242019”; to preset the time “13: 34”, input
“1334”.
The time is displayed in 24h format or 12h format, the date is displayed in British type, American
type or Chinese type, please set according to the requirement.
Brightness
Click
『
Brightness
』
to enter into display brightness setting interface. The brightness has 10 levels.
The equipment has the function of automatic brightness adjustment if external power supply is
unavailable. When there is no external power supply, power is supplied by battery, and is not