the infusion line is cut off or broken. Therefore, please periodically check the equipment it to avoid
above fault during the working period.
During infusion, periodically check the dripping state of the fluid and the fluid/drug in the intravenous
infusion bag/container, so as to ensure it is working correctly during infusion. This equipment doesn’t
directly measure the quantity of infusion fluid; therefore, it is possible that this equipment can’t detect
the free infusion flow under abnormal conditions. Even if the drop sensor is adopted, it is possible that
this equipment can’t detect the free infusion flow which is less than the specific value for the demands
of tolerance.
This equipment has the occlusion detection function for detecting and alarming when the infusion
needle deviates the position in the vein or the needle is not correctly punctured in the vein. However, it
only alarms when the occlusion pressure has reached certain numerical value. As a result, the
puncturing part may possibly have become reddish, swelling or bleeding. Additionally, it is possible
that the device doesn’t alarm for a long period if the actual occlusion pressure is lower than the alarm
threshold value, therefore, please periodically check the puncturing part. If there’s any abnormal
phenomenon for the puncturing part, please timely take suitable measures, such as puncturing again.
Only the infusion line, infusion needle and other medical components that meet the local laws and
regulations and the requirements covered in this User Manual can be adopted. It is suggested to use
the infusion lines with same brand as preinstalled in this equipment. Infusion accuracy cannot be
guaranteed if a infusion line brand that is not preinstalled is used. The drop sensor is based on infrared
sensor technology, if the drip sensor function is turned on, then the light-proof pipeline is not
applicable.
It is not allowed to disassemble, refit, or use this equipment for other purposes except normal infusion.
No one is allowed to repair this equipment except Triumph Medical Services LLC.
To avoid risk of electric shock,the equipment must only be connected to a supply mains with protectiv
e earth.