19
Set Infusion Parameters
Introduction to Infusion Parameters Setting
(1) The drug information can be displayed in the infusion running interface only when the drug
library is under active state.
(2) Click
『
Settings
』
icon in the main interface to enter sub-menu, find
『
Drug Library
』
menu
item, click to enter then select drug. To remove drug information from the running interface, go
back to Drug library and select “none”. Please refer to Drug Library on Page 21 for more info.
(3) For both the rate set in infusion parameter and the rate calculated by the system, the range is
the system default flow rate of the current working infusion line specification.
(4) If no VTBI (Volume to be infused) is Set then the fluid/drug in the infusion container will be
completely used.
Infusion Parameters Setting Range
Infusion Parameter
Parameter Range
VTBI
0-9999.99ml
Rate
0.1-1500ml/h
Time
1min-99hrs59min
Weight(Body weight)
0.1-300kg
Conc.unit(Concentration unit)
mcg/ml, mg/ml, g/ml, U/ml, kU/ml, IU/ml, EU/ml,
mmol/ml, mol/ml, kcal/ml
Volume(Fluid amount)
0.1-9999.99ml
Dose rate
0.1-9999.99
Dose unit
Unit(/kg)/min, Unit(/kg)/h,the Unit is Conc.unit
Drop rate
1-500 drops/min