30
button to stop alarm
Battery nearly empty
Low
1. When power is supplied only
with the built-in battery, under low
battery,
the
alarm
duration
is >30min
The
alarm
automatically
eliminates after connecting
the external power supply.
2. Battery ageing or the equipment
charging circuit is fault.
Please
contact
Triumph
Medical
Battery empty
High
When power is supplied by the
built-in battery only, under low
battery,
the
alarm
duration
is >30min
Immediately connect with
external power supply.
2. Battery ageing or the equipment
charging circuit is fault.
Please
contact
Triumph
Medical
Alarm
Alarm Level Reason
Solution
No battery
inserted
Low
Battery is removed
Keep
connecting
with
external
power
supply,
contact Triumph Medical
No power supply
Low
Under ON state, AC power supply
is adopted, but the AC power wire
is dropped during the process
The
alarm
automatically
eliminates after connecting
the external power supply.
No battery and No
power supply
High
Battery is removed and the AC
power wire is dropped
Connect
power
supply,
Contact Triumph Medical
Reminder alarm
Low
After installing infusion tube ,
under non-working or alarm state, it
is not operated within the set time
of the system
Click any button to stop
Standby time expired
Middle
During standby, after reaching the
standby time
Press
【
Stop
】
button to stop
alarm
KVO finished
High
KVO working time reaches 30min,
infusion pump stops working
Press
【
Stop
】
button to stop
alarm
Drop
sensor
connection
Low
When turning on the drop sensor,
the equipment is not connected with
the drop sensor
Connect the drop sensor, or
turn off the drop sensor in the
menu
Drop error
High
The angle of inclination of the drip
cup is too big or drop sensor is
installed lower than the drip cup
fluid level
Check the installation of drop
sensor or drip cup fluid level,
Press
【
Stop
】
button to stop
alarm
The specification of infusion line is
not accordant with the specification
displayed in the interface, which
causes drop rate error.
Check if the infusion line
specification
is
accordant
with displayed parameters, if
it is not accordant, it shall be