29
(2)
After disinfecting, wet the soft fabric with warm water, wring out excess fluid from fabric
and then wipe the equipment with it.
(3) Do not sterilize the equipment with a high-pressure steam sterilizer. Do not dry the
equipment with dryer or similar product.
Warning:
●
Please do not adopt Cidex OPA orthophthalaldehyde, methyl ethyl ketone or
similar solvent, otherwise, it may corrode the equipment.
Repair
Please contact Triumph Medical (888-388-3344 ext 3) to repair if there’s any fault or damage. Do
not disassemble and repair the equipment.
Alarm and Solution
Alarm Type
Alarm Level Reason
Solution
VTBI near end
Low
During infusion, the remaining time
reaches or is less than the set
nearing completion time
This
alarm
can’t
be
eliminated, and waits till
infusion completes
VTBI infused
High
The
preset
value
infusion
Completion
Press
【
Stop
】
button to stop
alarm
Pressure high
High
1. Line occlusion during infusion
Manually remove the reason
of occlusion, Press
【
Start
】
button to continue infusion
2. Fluid/drug in the actual infusion
line has high viscosity, but the
system occlusion level is set too
low
Rise the alarm Level, Press
【
Start
】
button to continue
infusion
3. The pressure sensor is damaged
Please
contact
Triumph
Medical
Pressure
near
threshold
Middle
Pipeline pressure increases close to
the preset blocking level.
Check the connection of the
pipeline, press
『
OK
』
button
to continue infusion
Pressure drop
Middle
When the pipeline pressure is high,
the pressure suddenly decreases.
Check the connection of the
infusion
pipeline,
press
『
OK
』
button to continue
infusion
Check upstream (if
applicable)
High
The upper part of the line is
obstructed during infusion, and in
turn drops the line pressure
intensity
Check if the rate regulating
adjuster or fluid stopping
device is opened at the upper
part of the line, Press
【
Stop
】