2.1 SpO
2
Monitoring
77
7
Sensors LNOP Blue Inf, LNOP TF-I, TC-I and LNCS TF-I, TC-I are not used for measurements in the presence of
movement.
8
Sensor LNOP YI is not used for measurements at low perfusion.
SENSORS
Use only Masimo oximetry sensors for SpO2 measurements.
The device provides operation with all sensors designed by Masimo for Masimo SET tech-
nology using the appropriate accessories (patient cables, adapters, connectors, etc.).
Before use, carefully read the LNOP
®
/ LNCS
®
sensor directions for use.
Tissue damage can be caused by incorrect application or use of an LNOP
®
/ LNCS
®
sensor,
for example by wrapping the sensor too tightly. Inspect the sensor site as directed in the
sensor Directions for Use to ensure skin integrity and correct positioning and adhesion of the
sensor.
Do not use damaged LNOP® / LNCS® sensors and patient cables. Do not immerse the
sensor in water, solvents, or cleaning solutions (the sensors and connectors are not water-
proof). Do not sterilize by irradiation, steam, or ethylene oxide. See the cleaning instructions
in the directions for use for reusable Masimo LNOP
®
/ LNCS
®
sensors.
When selecting an SpO2 sensor, make sure that the sensor used is appropriate for the pa-
tient category and overlap area. For more information on overlap areas, see the accompany-
ing operation instructions.
The tables (below) indicate some of the recommended sensor types, with the application ar-
ea and patient weight.
1. Reusable Sensors Masimo LNOP
®
(LNOP
®
patient cable is required)
No.
Sensor
Description
Application site
Weight range
1269
LNOP DC-I Reusable finger sensor
Finger
> 30 kg
1276
LNOP DC-IP Reusable pediatric finger sensor
Finger
10 - 50 kg
1794
LNOP TC-I Reusable ear sensor
Ear lobe
> 30 kg
1544
LNOP YI
Reusable multi-site sensor
Foot (hand, finger or
toe)
Toe
1 - 3 kg
> 3 kg
2. Disposable Adhesive Sensors Masimo LNOP
®
(LNOP
®
patient cable is required)
No.
Sensor
Description
Application site
Weight range
1001
LNOP Adt
Adult adhesive sensor
Finger
> 30
кг
1025
LNOP Pdt
Pediatric adhesive sensor
Finger
10 - 50 kg
1798
LNOP Neo-L
Neonate adhesive sensor
Foot
Finger
< 3 kg
> 40 kg
1800
LNOP Inf-L
Infant adhesive sensor
Great toe
3 - 20 kg
1651
LNOP
NeoPt-L
Sensitive skin neonate adhesive sen-
sor
Foot
< 1 kg
1970
LNOP Blue
Inf
Neonate, infant and pediatric sensor
for patients with congenital cyanotic
cardiac lesions
Great toe/ thumb
2.5 - 20 kg