3.1.3. Recording and Viewing Waveform Fragments
The device can store in memory up to 30 fragments of curves with duration of 10 s each (5 s
before the moment of recording and 5 s after). There are 2 modes for recording waveform frag-
ments:
manual - recording occurs when you press the "FREEZE" button;
automatic - recording occurs automatically when an arrhythmia is detected.
At the moment of recording fragments in the ECG window, a red "REC"
symbol appears for a short time.
1 - date and time of recording the fragment of curves displayed on the screen;
2 - button for selecting the serial number of the recorded fragment of curves;
3 - the reason for recording a fragment of the curves - manual mode or type of arrhythmia;
4 - number of the curve fragment displayed on the screen;
5 - values of parameters registered by the device at the moment of recording fragments;
6 - icon for starting / canceling printing on a thermal printer (displayed if there is a built-in thermal print-
er in the device);
7 - symbol for selecting the type of trend;
8 - close trend window;
9 - buttons for moving along the recorded fragments;
10 - field for selecting the type of the second curve;
11 - field for selecting the type of the third curve.
Figure 3-4. Window of curve fragments view
The scale of the ECG curves and the selection of the leads of the saved
fragment correspond to the settings in the device at the time of recording (6
curves are always memorized - three ECG curves, PPG, RSP, CAPNO). If
there is an IBP channel in the device, two curves of the IBP channel are ad-
ditionally stored.
The printing option is described in it. 3.1.4. An example of a printout of recorded fragments of