1.4 Alarms
64
1.4.9. Alarms of the Multigas Module
Alarm situation
Message (output
window)
Sound alarm
Light alarm
EtXXX value is less than bottom threshold
< 5 %
-
-
Flashing EtXXX
value
EtXXX value is less than bottom threshold
on
5…10 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
EtXXX value is less than bottom threshold
> 10 %
Three short
beeps every 5 s
The same on a
red background
EtXXX value is more than top threshold < 5
%
-
-
Flashing EtXXX
value
EtXXX value is more than top threshold on
5…10 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
EtXXX value is more than top threshold >
10 %
Three short
beeps every 5 s
The same on a
red background
FiXXX value is less than bottom threshold <
5 %
-
-
Flashing FiXXX
value
FiXXX value is less than bottom threshold
on
5…10 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
FiXXX value is less than bottom threshold >
10 %
Three short
beeps every 5 s
The same on a
red background
FiXXX value is more than top threshold < 5
%
-
-
Flashing FiXXX
value
FiXXX value is more than top threshold on
5…10 %
Two short beeps
every 5 s
The same on a
yellow back-
ground
FiXXX value is more than top threshold >
10 %
Three short
beeps every 5 s
The same on a
red background
No breathing
No Breathing!
APNEA!!!*
(multigas)
Continuous
signal
RSP = 0
Thermal stabilization of the measuring
chamber elements is in progress
Warming
(multigas)
-
-
Periodic zero autocalibration is in progress
Calibration
(multigas)
-
-
Multigas pneumatic path occlusion
Occlusion (multi-
gas)
-
-
Long-term multigas pneumatic path occlu-
sion
Long-term occlu-
sion (multigas)
-
-