3.11 Mainstream CO
2
Sensor
172
When returning to the factory settings, the alarm thresholds are set in accordance with it. 1.3.9,
and the values of the remaining parameters are set by default in accordance with the profile:
Profile
Units
Adult
Pediatric
Neonatal
Capnogram scale
mmHg
80
80
80
Capnogram sweep
speed
mm/sec
12.5
12.5
12.5
Apnea interval
sec
10
10
10
Averaging mode of
EtCO
2
readings
-
1 cycle
1 cycle
1 cycle
Balance gas type
-
air
air
air
CO
2
concentration
units
-
mmHg
mmHg
mmHg
3.11.3. Monitoring of CO
2
Concentration
The respiration curve (capnogram) is displayed in the capnometry channel window (see Fig. 3-
23), in the gas analysis parameters window - the measured values of EtCO
2
, FiCO
2
(see Fig. 3-
24), and in the respiration rate window (see Fig. 2-15) is the RR value measured by the capno-
gram (when the capnogram RR is selected as the source for calculating the RR in accordance
with it. 1.3.8).
1 - field for setting the scale of capnogram;
2 - field for setting the capnogram sweep speed;
3 - field for setting the interval of apnea;
4 - field for setting the mode of averaging readings by EtCO
2
;
5 - field for setting the type of breathing circuit;
6 - field for output capnogram and module messages;
7 - average physiological level of CO
2
concentration (5%);
8 - field for setting the type of output curve;
9 - real-time trend of EtCO
2
;
10 - zero setting button;
11 - field for setting the temperature of the breathing mixture;
12 - field for setting the balance gas;
13 - field for setting the real-time EtCO
2
trend.
Figure 3-23. Window of capnometry channel
1 - field for setting the top alarm threshold by EtCO
2
;
2 - field for sound alarm control for EtCO
2
;
3 - field for setting the bottom alarm threshold by EtCO
2
;
1
2
3
4
5
6
7