Sensor
169
3.11. MAINSTREAM CO
2
SENSOR
This module is supplied with the device as an option on a separate order and enables
to monitor:
concentration of
СО2 at the expiration end (EtСО2),
inspiratory CO2 concentration (FiCO2),
capnogram,
respiratory rate (RR), determined by capnogram.
The operation principle of the mainstream CO2 sensor is based on the method of measuring the
absorption of infrared radiation in the absorption spectrum of carbon dioxide.
Only mainstream CO
2
modules supplied by Triton Electronic Systems Ltd.
are compatible with the monitor.
The external mainstream CO2 sensor is connected to the device through the "Infoport" con-
nector installed on the left side panel of the device (Fig. 1-4, position 10).
The module is connected to the ventilation circuit using matching devices - adapters of sampling
lines (ventilation adapters) of various types, which enable monitoring patients from newborns to
adults.
1 - sensor housing;
2 - ventilation adapter;
3 - connector for an information port;
4 - sensor cabel.
Figure 3-21. External mainstream CO
2
sensor
Attention! It is not recommended to connect the external mainstream CO2
sensor when the device is turned on.
4
2
1
3