Annex 6. Electromagnetic Environment
192
Annex 3. Calibration of the Touch Panel
Calibration of the touch panel is intended for the correct positioning of the touch buttons on the
touch panel in accordance with their graphical image. This calibration may be required as a re-
sult of the electrical parameters drift of the touch panel during long-term operation of the device.
To calibrate the touch panel of the device,
before turning it on
, press the "FREEZE" button and
do not release it until the calibration window appears on the device screen with the message:
«Touchscreen calibration: touch the screen in the specified point»
In the lower left corner, the symbol “ ” indicates the first point suggested for pressing. Press in
the center of the symbol with the greatest possible accuracy (for example, with a pencil).
Perform the same procedure for the second point.
Then you will the third point marked with the symbol This point is intended to check the
proper calibration. If the calibration is performed correctly, then after pressing the third point, the
device eautomatically switches to operating mode. If errors occur during the calibration process,
the device returns to the beginning of the calibration process.
Exit from the calibration mode is possible only if the calibration is complet-
ed successfully or the device is turned off.