TRC ELECTRONICS, INC.
Providing exceptional customer service since 1982
TRC ELECTRONICS, INC.
1.888.612.9514
www.
trcelectronics
.com
Designers Manual
Advanced Energy CoolX600 Series
TTaabbllee 11‑‑11.. CoolX galvanic isolation barriers (Continued)
IIssoollaattiioonn BBaarrrriieerr
TTyyppee
W
Wiitthhssttaanndd VVoollttaaggee
Output to output (one
module to another)
Basic (1 x MOPP)
1850 VAC
Output to output (within a
dual module)
Functional
500 VDC
LLO
OW
W VVO
OLLTTAAG
GEE D
DIIR
REEC
CTTIIVVEE ((LLVVD
D)) 22000066//9955//EEC
C
The LVD applies to equipment with an AC input voltage from 50 V to 1000 V or a
DC input voltage from 75 V to 1500 V. The CoolX series is CE marked to show
compliance with the LVD.
The relevant European standard for the CX06S model is EN 60950-1 2nd Edition
(information technology) and 62368-1 Edition 2. The relevant European standard for
the CS06M model is EN60601-1 3rd Edition (Medical Devices Directive).
The full table of safety certifications follows.
TTaabbllee 11‑‑22.. CoolPac safety certifications
C
CoooollPPaacc C
Chhaassssiiss
SSaaffeettyy SSttaannddaarrddss
AAddddiittiioonnaall D
Deettaaiillss
CX06S
IEC/EN 60950-1 Edition 2
and all national deviations
UL 60950-1/CSA C22.2
No 60950-1 Edition 2
5000 m (16,404′) altitude,
100 VAC to 240 VAC ±
10%
IEC 62368-1 Edition 2
IEC 62368-1 (2014)
Edition 2
5000 m (16,404′) altitude,
100 VAC to 240 VAC ±
10%
CX06M
IEC/EN 60601-1 Edition 3
and all national deviations
IEC 60601-1(2005), EN
60601-1(2006) ANSI/
AAMI ES 60601-1(2005)
CAN/CSA C22.2 No.
60601-1 (2008)
5000 m (16,404′) altitude,
100 VAC to 240 VAC ±
10%
Advanced Energy
®
CoolX
®
600 Modular Power Supply
57010208-00B
Safety and Product Compliance Guidelines
1‑4