APPENDIX C: DISPLAYED ERROR CODES
CODE
MODE
MEANING / POSSIBLE SOLUTION
Normal Operating
Mode
Gross Overload. A weight greater than the rated capacity has been applied
to the scale. Remove the weight from the platter or try re-calibrating the
scale. Otherwise, check for a bad load cell connection or possible load cell
damage due to overloading.
Err 0
Span Calibration
Mode (F17)
Keyed-in weight value is larger than full-scale capacity. Use a smaller test
weight or check keyed-in value.
Err 1
Span Calibration
Mode (F17)
Keyed-in weight value is less than 1% of full-scale capacity. Use a larger
test weight or check keyed-in value.
Err 2
Span Calibration
Mode (F17)
There is not enough load cell signal to produce the internal counts
necessary to properly calibrate the scale. First check all load connections.
Use F16 mode to view internal counts.
Err 3
All Modes
Non-volatile memory read error. One or more setup parameters have been
lost.
Err 4
All Modes
Non-volatile memory write error. Indicator needs service.
Err 9
Normal Operating
Mode
Span calibration value has been lost. Re-calibrate the scale.
Slowly
Flashing
Normal Operating
Mode
Indicates that the battery voltage is becoming too low for normal operation.
Batteries should be replaced soon.
Quickly
Flashing
Normal Operating
Mode
Indicates that the battery voltage has become too low for normal operation.
Batteries should be replaced now.
Flashes
“bAtt”
Normal Operating
Mode
Indicates that the battery voltage has become too low for normal operation;
indicator will shut off automatically after 20 seconds. Batteries should be
replaced now.